学術雑誌論文 An application of CCD read-out technique to neutron distribution measurement using the self-activation method with a CsI scintillator plate

Nohtomi, Akihiro  ,  Kurihara, Ryosuke  ,  Kinoshita, Hiroyuki  ,  Honda, Soichiro  ,  Tokunaga, Masaaki  ,  Uno, Heita  ,  Shinsho, Kiyomitsu  ,  Wakabayashi, Genichiro  ,  古場, 裕介  ,  Fukunaga, Junichi  ,  Umezu, Yoshiyuki  ,  Nakamura, Yasuhiko  ,  Ohga, Saiji

内容記述
In our previous paper, the self-activation of an NaI scintillator had been successfully utilized for detecting photo-neutrons around a high-energy X-ray radiotherapy machine; individual optical pulses from the self-activated scintillator are read-out by photo sensors such as a photomultiplier tube (PMT). In the present work, preliminary observations have been performed in order to apply a direct CCD read-out technique to the self-activation method with a CsI scintillator plate using a Pu-Be source and a 10-MV linac. In conclusion, it has been revealed that the CCD read-out technique is applicable to neutron measurement around a high-energy X-ray radiotherapy machine with the self-activation of a CsI plate. Such application may provide a possibility of novel method for simple neutron dose-distribution measurement.