紀要論文 コーンビーム CT 画像の散乱線補正が領域分割精度に 与える影響の検討

久保田, 純加

582017-03-31 , 法政大学大学院理工学・工学研究科
ISSN:21879923
NII書誌ID(NCID):AA12677220
内容記述
The purpose of the study is to use a cone beam CT for the radiation therapy. A cone beam CT system attached to a linear accelerator has no collimator to avoid scattered photons originated in a patient body.So the reconstructed CT values are affected by scattered photons, as a result of it, the CT values are inaccurate. To remove the scattered photons, a scatter kernel method was used in the process of image reconstruction with a maximum likelihood expectation maximization method. The feasibility of the proposed method was evaluated with the result of segmented regions comparing the multi-slice CT and scatter corrected cone beam CT image. Our proposed method could segment the bone, soft tissue and lungareas as well as with the result of the multi-slice CT images.Key Words : cone beam CT, maximum likelihood expectation maximization method, scatter correction