Journal Article Time-resolved dose evaluation in an X- and gamma-ray irradiated silver-activated glass detector for three-dimensional imaging applications

Kurobori, Toshio  ,  Itoi, H.  ,  Yanagida, Y.  ,  Chen, Y.Q.

Description
Ag-activated phosphate glass based on the radiophotoluminescence (RPL) phenomenon has been used as the most commonly known RPL material and as an accumulated-type passive detector. In this work, the transient-state evaluation of the dose distributions achieved by X- and gamma-ray irradiations within the Ag-activated phosphate glass was performed using a time-resolved technique for the first time. Specifically, the blue RPL intensity ascribed to the electron-trapped Ag0 centres as a function of the depth at the vicinity of the surface was investigated for different types of radiation and a wide range of energies. In addition, the dose distributions at each layer within the glass confirmed by the time-resolved measurement were compared with those reconstructed by a disk-type transparent glass detector based on the blue RPL with a diameter of 100 mm. © 2015 Elsevier B.V. All rights reserved.
Full-Text

http://dspace.lib.kanazawa-u.ac.jp/dspace/bitstream/2297/42994/1/ED-PR-KUROBORI-T-6.pdf

Number of accesses :  

Other information