Departmental Bulletin Paper Effects of Neo Automatic Code on the Accuracy of Chest Compression Depths in Cardiac Arrest Patients

Darmawan, Rendi Editya  ,  Sujianto, Untung  ,  Rochana, Nana

67pp.161 - 165 , 2018-05 , Hiroshima University Medical Press
ISSN:0018-2052
NCID:AA00664312
Description
Objective: to analyze the effects of manual chest compression using neo automatic code on the accuracy of compression depths. Methods: This study used a quantitative, post test quasi-experimental design with a control group. The samples were 74 cardiac arrest patients in two hospitals in Surakarta and Klaten, who were selected using purposive sampling technique. The data were analyzed by the Mann-Whitney test. Results: the mean of accuracy of compression depth in the control group was 68.10 ± 17.60%, and in the treatment group was 83 ± 6.04%. The result of statistical analysis showed that there were differences in the accuracy of compression depth in the intervention and control group with p-value of 0.000. Conclusion: there were effects of manual chest compression using neo automatic code on the accuracy of compression depths. Neo automatic code could improve the accuracy of chest compression depths.
Full-Text

http://ir.lib.hiroshima-u.ac.jp/files/public/4/45860/20180619110338247329/HiroshimaJMedSci_67s_161.pdf

Number of accesses :  

Other information