学術雑誌論文 Accuracy assessment methods of tissue marker clip placement after 11-gauge vacuum-assisted stereotactic breast biopsy: comparison of measurements using direct and conventional methods

Yatake, Hidetoshi  ,  Sawai, Yuka  ,  Nishi, Toshio  ,  Nakano, Yoshiaki  ,  Nishimae, Ayaka  ,  Katsuda, Toshizo  ,  Yabunaka, Koichi  ,  Takeda, Yoshihiro  ,  Inaji, Hideo

24 ( 4 )  , pp.593 - 600 , 2017-07 , Japanese Breast Cancer Society
ISSN:1340-6868
NII書誌ID(NCID):AA1103354X
内容記述
BACKGROUND: The objective of the study was to compare direct measurement with a conventional method for evaluation of clip placement in stereotactic vacuum-assisted breast biopsy (ST-VAB) and to evaluate the accuracy of clip placement using the direct method. METHODS: Accuracy of clip placement was assessed by measuring the distance from a residual calcification of a targeted calcification clustered to a clip on a mammogram after ST-VAB. Distances in the craniocaudal (CC) and mediolateral oblique (MLO) views were measured in 28 subjects with mammograms recorded twice or more after ST-VAB. The difference in the distance between the first and second measurements was defined as the reproducibility and was compared with that from a conventional method using a mask system with overlap of transparent film on the mammogram. The 3D clip-to-calcification distance was measured using the direct method in 71 subjects. RESULTS: The reproducibility of the direct method was higher than that of the conventional method in CC and MLO views (P = 0.002, P < 0.001). The median 3D clip-to-calcification distance was 2.8 mm, with an interquartile range of 2.0-4.8 mm and a range of 1.1-36.3 mm. CONCLUSION: The direct method used in this study was more accurate than the conventional method, and gave a median 3D distance of 2.8 mm between the calcification and clip.

このアイテムのアクセス数:  回

その他の情報