Journal Article Deans Switch型SilFlowを用いた血中ダイオキシン類測定におけるソルベントカット大量注入法の開発
Development of Solvent-Cut Large-Volume Injection System using Deans Switch Type SilFlow for Dioxin Determinations in Blood

安武, 大輔  ,  飛石, 和大  ,  平川, 博仙  ,  新谷, 依子  ,  小木曽, 俊孝  ,  堀, 就英  ,  梶原, 淳睦  ,  香月, 進  ,  三苫, 千景  ,  古江, 増隆

108 ( 3 )  , pp.102 - 110 , 2017-03-25 , 福岡医学会
ISSN:0016-254X
NCID:AN00215478
Description
We developed solvent-cut large-volume injection system (SCLV) using Deans switch type SilFlow in dioxins measurement of human blood. The SilFlow worked in the same mechanism as the conventional SCLV for dioxins measurement, showed high reproducibility in dioxins measurement of the standard solution and the control sample. Since the SilFlow has very small dead volume as compared with the conventional method, the influence of contamination is small, and the handling is easy. In this study, we proved that SilFlow is available for an alternative to SCLV in blood dioxins analysis.
Full-Text

https://catalog.lib.kyushu-u.ac.jp/opac_download_md/1810707/p102.pdf

Number of accesses :  

Other information